HC350S, HCINT350S Series Power Module Application Guidelines
SAVE THESE INSTRUCTIONS – This manual contains IMPORTANT SAFETY
INSTRUCTIONS for that should be followed during installation and maintenance of the
UPS and batteries.
The HC350S and HCINT350S series of hospital cart power modules are designed to
provide AC voltage output power from either utility provided AC input voltage or from
inverting 12VDC power from a sealed lead acid battery. To ensure proper operation the
following guidelines should be observed:
1. The power module can support AC loads up to a maximum of 400VA or 300W.
2. The power module charger is intended to be used with a 12V sealed lead acid
absorbed glass mat (AGM) battery with a 33Ah to 66Ah rating. Batteries rated at
less than 33AH are not recommended since the charger current of 12A would
reduce the battery cycle life of smaller batteries.
3. The appliance inlet dongle serves as a disconnect device and shall be easily
accessible.
4. Alternately, the socket-outlet of the end-product shall be installed near the
equipment and shall be easily accessible or disconnect device provided within
the end-product.
5. Adequate provision for airflow needs to be provided at both the inlet and outlet
vents of the power module. There should be at least 0.5 inches of clear space on
the fan end and on the vented end of the power module. A baffle or air deflector
is recommended at the fan end so that the air intake is raised at least 6 inches
above floor level to prevent dust accumulation inside the power module.
6. The power module may be mounted in any orientation. Four 8-32 threaded
inserts are provided on the surface with the AC input cord to facilitate mounting.
7. The AC voltage input and output connections are made with user supplied cables
to the IEC input and output connectors on the power module. Use only with
approved plug/cord sets according to the equipment electrical ratings. No
external fusing is needed since safe operation is assured by internal non-
replaceable 10A fuses. However, if a customer wishes to prevent the internal
fuses from blowing in the event of an extreme overload, then an external fuse or
circuit breaker rated at 6A to 10A (120VAC models) or 3A to 5A (230V models)
should be provided on the AC voltage input.
8. The 12V battery input is made with a user supplied cable to the Anderson DC
connector on the power module. Proper polarity must be observed. The
recommended wire size is 8AWG in order to keep the voltage drop in the cable to
less than 0.1V. The length of cable between the battery and the power module
should be less than 30 inches. A user supplied 40A to 50A automotive style fuse
is required in the positive lead of the battery (within 18 inches of the battery
positive terminal).
23-08-312-932965 REV E
Series AGSM5396, AGTR0829
1000 Eaton Boulevard
Cleveland, OH 44122
United States
TrippLite.Eaton.com
Tripp Lite is now part of Eaton.

Advertisement

loading

Summary of Contents for Tripp Lite HC350S Series

  • Page 1 30 inches. A user supplied 40A to 50A automotive style fuse is required in the positive lead of the battery (within 18 inches of the battery positive terminal). 1000 Eaton Boulevard Cleveland, OH 44122 23-08-312-932965 REV E United States TrippLite.Eaton.com Tripp Lite is now part of Eaton.
  • Page 2 COMM2 jacks should be connected together. If the cables are not connected or if they are interchanged, the power module will not operate. The RUI may be either a Tripp Lite RUI or a customer designed RUI that meets Tripp Lite specifications.
  • Page 3 Guías de Aplicación para los Módulos de Potencia de las Series HC350S y HCINT350S Series AGSM5396, AGTR0829 GUARDE ESTAS INSTRUCCIONES – Este manual contiene INSTRUCCIONES DE SEGURIDAD IMPORTANTES que deben seguirse durante la instalación y el mantenimiento del UPS y las baterías. Los módulos de potencia para carros de hospital de las series HC350S y HCINT350S están diseñados para proporcionar una potencia de salida de voltaje de CA suministrada por la red pública o a partir de la inversión de potencia de 12VCD de una...
  • Page 4 COMM2 también deben conectarse juntos. Si los cables no están conectados o si están intercambiados, el módulo de potencia no funcionará. El RUI puede ser un RUI de Tripp Lite o un RUI diseñado por el cliente que cumpla con las especificaciones de Tripp Lite.
  • Page 5 Detalles técnicos Tipo de sistema de suministro eléctrico: TN Protección contra la penetración: IP20 Entorno de Almacenamiento: 0% a 95% de HR, Sin Condensación, 0 ºC ~ 50 ºC [32 ºF ~ 122°F] . Rango de temperatura de operación ambiente: 0 ºC ~ 40 ºC [32 ºF ~ 104 ºF] Estándares relevantes para fabricación, prueba o uso: IEC/EN 62040-1, IEC 60601-1 23-08-312-932965 REV E...
  • Page 6 HC350S, HCINT350S-Baureihe Leistungsmodul - Applikationsrichtlinien Baureihe AGSM5396, AGTR0829 BEWAHREN SIE DIESE ANWEISUNGEN AUF – Dieses Handbuch enthält WICHTIGE SICHERHEITSHINWEISE, die bei der Installation und Wartung der USV und Batterien beachtet werden sollten. Die Leistungsmodule der HC350S- und HCINT350S-Baureihe für Krankenhauswagen sind so konzipiert, dass die Ausgangsleistung der AC-Spannung entweder von der bereitgestellten AC-Eingangsspannung des Versorgungsunternehmens oder durch die Invertierung der 12-VDC-Leistung einer versiegelten Bleisäure-Batterie geliefert wird.
  • Page 7 COMM2-Buchsen sollten jeweils miteinander verbunden werden. Wenn die Kabel nicht angeschlossen sind oder ausgetauscht werden, funktioniert das Stromversorgungsmodul nicht. Die RUI kann entweder eine RUI von Tripp Lite oder ein vom Kunden entwickeltes RUI sein, das die Spezifikationen von Tripp Lite erfüllt.
  • Page 8 15. ACHTUNG: Zweipolig/Neutral Absicherung. Technische Daten Art des elektrischen Versorgungssystems: TN Schutz vor Eindringen: IP20 Speicherumgebung: 0 bis 95 % RF, nicht kondensierend, 0 bis 50 ºC. Umgebungsbetriebstemperaturbereich: 0 bis 40 °C Einschlägige Normen für die Herstellung, Prüfung oder Verwendung: IEC/EN 62040- 1, IEC 60601-1 1000 Eaton Boulevard Cleveland, OH 44122...

This manual is also suitable for:

Hcint350s series