See the Diagnostic Cardiology Supplies and Accessories Guide, 2102946-001.
Storage
Follow the storage instructions:
• Store the device, trolley, and acquisition module in a clean and dry, well-ventilated
area protected from dust, moisture, insects, vermin, and extremes of temperature
and humidity.
• Use below guidelines when storing acquisition modules:
• Hang the device using a holder, if available.
• If cables and leadwires are attached, route the additional length cables and
store them in the basket behind the device.
Other Cleaning and Disinfection Agents
Super Sani-Cloth
which has been validated on the device. However, below products are compatible
with the device and may be used for cleaning and disinfection. It should be noted that
these products have not been validated and their efficacy cannot be guaranteed.
• PDI Easy Screen Cleaning
• PDI Super Sani-Cloth
• PDI Sani-Cloth
• Clinell Sporicidal Wipes
• PDI Sani-Cloth
• PDI Sani-Cloth
• PDI Sani-Cloth
• Clorox Healthcare
• Oxivir
• Clinell Universal Range
• Cleanisept Wipes
• Mikrozid Sensitive Wipes
• Caviwipes
• Phenol 2% (v/v)
• Ethanol (ethyl alcohol) 96% (v/v)
• Hydrogen Peroxide 20% (v/v)
• Sodium Hypochlorite (NaOCl) 5% solution)
• Isopropyl alcohol 70% (m/m)
2088531-370-2
®
Wipes is the recommended cleaning and disinfection solution,
®
®
®
Bleach
®
AF3
®
Plus
®
HB
®
Hydrogen Peroxide Cleaner Disinfectant Wipes
®
Tb Wipes
MAC VU360
™
Resting ECG Analysis System
Cleaning and Disinfection
119