Miele PG 8583 WW ADP LD Operating Instructions Manual page 16

Washer-disinfector for laboratory glassware and laboratory utensils
Table of Contents

Advertisement

Warning and Safety instructions
16
Cleaning agents containing chlorine can damage the elastomers of
the machine.
If the use of cleaning agents containing chlorine is required, a
maximum temperature of 75 °C in the "Main wash" programme
phases is recommended (see programme chart).
Cleaning agents containing chlorine must not be used in machines
supplied (ex works) with special oil-resistant elastomers for oil and
grease applications.
Pre-treatments with cleaning or disinfecting agents can create
foam, as can certain types of soiling and chemical agents. Foam can
have an adverse effect on the cleaning and disinfection result.
Processes must be set up such that foam cannot escape from the
wash cabinet. It would hinder the correct functioning of the machine.
The process used must be monitored on a regular basis by the
supervisor to check foaming levels.
To avoid the risk of damage to the machine and any accessories
used with it caused by chemical agents, soiling and any reaction
between the two please read the notes in "Chemical processes and
technology".
Where a chemical agent is recommended on technical application
grounds (e.g. a cleaning agent), this does not imply that the
manufacturer of the machine accepts liability for the effect of the
chemical on the items being cleaned.
Please be aware that changes in formulation, storage conditions etc.
which may not be publicised by the chemical manufacturer, can have
a negative effect on the cleaning result.
When using process chemicals, always consult the instructions
issued by individual manufacturers. Process chemicals must only be
used for the purpose they are designed for by the manufacturer to
avoid any material damage or the occurrence of very strong chemical
reactions (e.g. oxyhydrogen explosion).
Always follow the relevant manufacturer's instructions on storage
and disposal of chemical agents.
In critical applications where very stringent requirements have to
be met, it is strongly recommended that all the relevant factors for the
process, such as chemical agents, water quality etc. are discussed
with the Miele Application Technology specialists.
If the cleaning result is subject to particularly stringent
requirements, e.g. in chemical analysis, the operator must carry out
regular quality control to ensure that required standards of cleanliness
are being achieved.
The mobile units, baskets, modules and inserts that hold the load
must be used only as intended.
Hollow items must be thoroughly cleaned, internally and externally.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pg 8583 ww ad ld62858306gb62858326gb62858304gb62858305za

Table of Contents