Miele PG 8583 WW ADP LD Operating Instructions Manual page 134

Washer-disinfector for laboratory glassware and laboratory utensils
Table of Contents

Advertisement

Problem solving guide
Problem
The cleaning result is un-
satisfactory.
Items made of glass are
showing signs of corro-
sion.
134
Cause and remedy
Mobile units, baskets, modules and inserts were not
suitable for the load.
Select mobile units, baskets, modules and inserts which
are suitable for the task.
Mobile units, baskets, inserts and modules were incorrectly
loaded or overloaded.
Arrange the wash load correctly according to the
information in the Operating instructions.
Avoid overloading the mobile units, baskets, modules
and inserts.
The programme was not suitable for the soiling.
Select a suitable programme.
or
Adjust the parameters to suit the task.
A spray arm is blocked.
Ensure the spray arms are not obstructed when
arranging the wash load.
Injector nozzles in the mobile units, baskets, modules or
spray arms are blocked.
Check the nozzles and clean them as necessary.
The filters in the wash cabinet are dirty.
Check the filters and clean them if necessary.
Mobile units, baskets or modules were not correctly fitted
to the water connection.
Check the adapter.
The items are not suitable for machine reprocessing.
Only use items which are declared by their manufacturer
as suitable for machine reprocessing.
Neutralisation has not taken place during the programme.
Check the level in the supply container and vent the
dispensing system if necessary.
The wash temperature was too high.
Select a different programme.
or
Reduce the wash temperature.
Cleaning agents used were too alkaline.
Use a milder cleaning agent.
or
Reduce the concentration of the cleaning agent.