Fill dispensing system
The
pensing system. The function must be carried out after a container for
process chemicals is filled or replaced with a full container.
The function can only be carried out if there are no load carriers in
the wash cabinet.
Press the
Fill dispensing system
Press the button for the required dispensing system, e.g.,
The individual steps are shown on the display during the process.
Once the process has finished, a message will appear on the display:
Dispensing system DOS 1 filled.
or
Dispensing system DOS 1 not filled. Repeat Fill dispensing system
Press OK to confirm the message.
If the process of filling a dispensing system is not repeated suc-
cessfully, contact Miele Technical Service.
If the machine is equipped with a drying unit, the coarse and fine fil-
ters in the drying unit must be replaced by Miele Technical Service on
a regular basis.
You can read off the operating times and the times left for the individ-
ual filters in the

Filter maintenance

Set up dispensing system
This must be performed if:
- The connected process chemical has been changed (different vis-
cosity).
- A new dispenser pump or a new flow sensor has been installed
- The dispensing system needs to be checked, e.g., in the event of
qualification, requalification, or a periodic check.
During setup, the dispenser pump is activated for a defined number
of pulses. In parallel to this, the control module counts the pulses
emitted by the flow sensor. After the measured volume has been en-
tered into the control module, the module calculates the conversion
factors for the throughput of the dispenser pump in µl/Imp and the
pulse rate of the flow sensor in µl/Imp.
 function is used to prime the relevant dis-
 button.
 menu.
is used to calibrate the dispensing systems.
DOS 1
.
.