1.2 Installation Conditions

1.2 INSTALLATION CONDITIONS

1.2.1 Power Requirements

1.2.2 Installation Place

1.3 INSTALLATION ENVIRONMENT

1.3.1 Operating Temperature Range

1.3.2 Operating Humidity Range

1.3.3 Storage Temperature Range

Check that the following conditions are satisfied for the installation
of the spectrophotometer.
Power supply
: AC 100, 115, 220, 230 or 240 V
Fluctuation should be within 10% of rated
voltage
voltage.
Frequency
: 50 or 60 Hz
Fluctuation should be within 0.5Hz of rated
frequency.
Power source : 300 VA or more
capacity
It is recommended to provide a capacity of 500
VA or more in consideration of joint use of
associated accessory devices.
Grounding line : Grounding resistance should be 100  or less.
Floor area
Work table strength : Flat-top table capable of withstanding
5 to 35 C
In order to perform measurement under the most stable condition,
we recommend that the instrument be used in an air-conditioned
room at 20 to 25 C.
45 to 85%
-20 to +60 C
1 - 2
: Width of 500 mm and depth of 605 mm or
more. Give space at least 200 mm on
both sides of the main unit.
weight of 29 kg or more.