Control Panel - Buffalo FS130 Instruction Manual

Uvc sterilising cabinets
Table of Contents

Advertisement

Operating
This appliance is intended for professional use only and should be operated by fully trained and qualified
personnel.

Control Panel

Digital timer:
To start sterilisation cycle and countdown timer
from 30 minuites press START
START/STOP button:
Press the button to start sterilizing /stop sterilizing
1. Connect the appliance to a mains power supply.
2. Open the door, place items that require sterilizing in and close the door.
3. Press START and the 30 min cycle will begin.
4. For the first 20 seconds of the cycle the door can be opened, after this the red light will illuminate
and the door must remain shut until the end of the cycle or the cycle is stopped.
5. To stop the cycle before 30 minuites press the STOP button. The timer will count down from 5
minuites to eliminate ozone in the unit and the door must remain shut. When the green light
illuminates the door can be opened.
6. When a cycle is complete the timer will default back to 00:00.
When the green light is illuminated
it is safe to open the door
When the red light is illuminated
the door must not be opened