M20011 REL-0000 (2103)
Media compatibility
Pressure Types
DN:
Differential pressure, non-isolated sensors for use with
▪
clean, dry, non-corrosive gases only.
CN:
Compound non-isolated sensors for use with clean, dry,
▪
non-corrosive gases only.
AN:
▪
Absolute non-isolated sensors for use with clean, dry,
non-corrosive gases only.
DI:
Differential pressure, isolated sensors for use with gases
▪
and liquids compatible with 316LSS and Viton O-rings
CI:
Compound or Absolute pressure sensors for use with
▪
gases and liquids compatible with 316LSS.
AI:
Compound or Absolute pressure sensors for use with
▪
gases and liquids compatible with 316LSS.
Pressure Limits
▪
DN units:
CN units:
▪
AN units:
▪
DI units:
▪
(LO) only is 3x range or 150 PSI, whichever is less.
CI units:
▪
AI units:
▪
User Manual for Smart Manometer
3x range or 200 psi, whichever is less.
3x range or 200 psi, whichever is less.
3x range or 200 psi, whichever is less.
1000 psi common mode, P1 (HI) only is 3x range, P2
2x range.
2x range.
58 of 60