Test Results Form - GE DINAMAP ProCare Service Manual

Table of Contents

Advertisement

Test Results Form

LEAKAGE
Leakage Result (mmHg)
PRESSURE TRANSDUCER VERIFICATION
Pressure reading at 200mmHg, top display - Systolic
Pressure reading at 200mmHg, bottom display -
Diastolic
Pressure reading at 150mmHg, top display - Systolic
Pressure reading at 150mmHg, bottom display -
Diastolic
Pressure reading at 100mmHg, top display - Systolic
Pressure reading at 100mmHg, bottom display -
Diastolic
Pressure reading at 50mmHg, top display - Systolic
Pressure reading at 50mmHg, bottom display - Diastolic
OVERPRESSURE VERIFICATION
Overpressure threshold, Adult (mmHg)
Overpressure threshold, Neonate (mmHg)
BUTTONS
BP Determination Initiated
"E80" displayed on SYSTOLIC display
Audible alarm can be silenced
"Silenced" LEDs flash
Overpressure alarm can be cleared
Alarm button is functioning
DISPLAY
All 7-Segment LEDs Light Correct Color
All Discrete LEDs Light, Correct Color
EXTERNAL DC DETECTION
Charging indicator LED illuminated
BP DETERMINATION
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Heart rate reading (bpm)
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Revision C
Maintenance: Test Results Form
Description
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
Min
Max
Actual
0
6
197
203
197
203
147
153
147
153
97
103
97
103
47
53
47
53
305
325
150
165
Pass-Fail-N/A
4-17