Dio Data Type/Format; Data Type; Factory Calibrate; Sensor Writeable - Omega SP-005 User Manual

Thermocouple and rtd smart probe
Table of Contents

Advertisement

5.3.1.2

DIO Data Type/Format

7

Smart Sensor

Sensor Writable
0
Note:
Please refer to the Smart Sensor Interface Technical Guide for more information regarding this descriptor.
5.3.1.3

Data Type

The 4-bit Data Type field determines the type of date of the specific sensor.
5.3.1.4

Factory Calibrate

The Factory Calibrate bit is not used for DIO types.
5.3.1.5

Sensor Writeable

If the Sensor Writeable bit is set the sensor value may be overwritten with a preset value. This capability is useful in
sensors such as up/down counters, where a preset, or possibly a zero value must be written to the sensor value.
5.3.1.6
Smart Sensor
Refer to the Smart Sensor Device Interface documentation.
5.3.2

DIO Input Configuration

7

Available

Assigned

0
5.3.2.1

Lock

If set, the user specified units of measure string (4-character maximum) will be used in place of the default DIN.
5.3.2.2

Apply Scaling

If set, the user defined Offset and Gain values will be used to adjust the sensor reading:
Result = (Raw Reading * Gain) + Offset
5.3.2.3
Assigned
The Assigned bit will always read as 0. Refer to the Smart Sensor Device Interface documentation for further
information.
5.3.2.4
Available
The Available bit will always read as 0. Refer to the Smart Sensor Device Interface documentation for further
information.
6
5
Factory Calibrate
0
0
DIO Input Configuration
6
5
Apply
Scaling
0
1
DIO Data Type
4
reserved
0
4
3
Lock
Sub Channel Selection
?
0x03 == bits 0 and 1
3
2
Data Type
6 == Floating point
2
1
1
0
0
19 |
P a g e