4.3.1

Cold Junction Calibration

The SP-005 has automatic Cold Junction Compensation and is factory calibrated so that in most cases it needs no
adjustment. If increased accuracy is desired, Cold Junction Calibration can be performed as described below:
Step 1:
Ensure your thermocouple has been configured in the previous section and click Calibration beneath the
input interface.
Step 2:
Insert the probe into 0°C (32°F) reference as outlined in SYNC and allow it to stabilize. Once the
thermocouple is stable in a 0°C environment, click Calibrate.
For thermocouples that do not read to 0°C, One-Point Calibration can be used instead of Ice-Point Calibration.