Miele PG 8583 CD Operating Instructions Manual page 90

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

Further settings
Resetting the
counter
Setting the interval
90
The counter for the cleaning interval may be reset only after
cleaning has been completed.
Filter combination
Reset
-
The counter is reset.
Interval
-
Select an option using the  and  arrow buttons and confirm your
choice with OK.
The interval depends on the number of programme sequences and
must be set on the basis of usage and the expected number of
particles/solids in the soiling.
Example:
For weekly cleaning with 2 programme sequences per day and 5
working days in the week, this yields an interval of 10 (2 x 5 = 10).
With a higher incidence of particles, a shorter interval should be
selected in order to clean the filters several times weekly.
The setting value is entered in increments of 5. The possible range is
shown in the bottom line of the display.
Use the arrow buttons  (higher) and  (lower) to set the interval.
Press OK to save the setting.
Reset
Interval
Interval
10
(5 - 100)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pg 8583 cd ww adp cmPg 8583 cd ww ad cm oil

Table of Contents