Miele PG 8583 CD Operating Instructions Manual page 43

Washer-disinfector for laboratory glassware and laboratory utensils
Hide thumbs Also See for PG 8583 CD:
Table of Contents

Advertisement

...wide necked
...narrow necked
Wash load...
Wash load items with wide necks, e.g. beakers, wide necked
Erlenmeyer flasks and petri dishes, or cylindrical items, e.g. test
tubes, can be cleaned inside and out by rotating spray arms.
To do this the wash load is positioned in full, half or quarter inserts
and placed in an empty lower basket or an upper basket with a spray
arm.
Baskets with special injector modules are available for wash load
items with narrow necks, e.g. narrow necked Erlenmeyer flasks,
round bottomed flasks, measuring flasks and pipettes.
The injector units and modules come with their own operating
instructions.
When loading please note:
- Place petri dishes in the appropriate insert with the dirty side facing
towards the middle.
- Place pipettes with the pointed end facing downwards.
- Quarter segment inserts should be positioned at a minimum 3 cm
distance from the edge of the upper or lower basket.
- Position quarter segment inserts for test tubes around the middle to
leave the corners of the upper or lower basket free.
- Use a cover net to avoid breakages if required.
Areas of application
43