Flowmeter Sensor Model Fcb300 / Fch300; Temperature Class - ABB CoriolisMaster FCB330 Commissioning Instruction

Coriolis mass flowmeter
Table of Contents

Advertisement

8.4

Flowmeter sensor model FCB300 / FCH300

8.4.1

Temperature class

Model FCx3xx-F1..., in Class I Div. 1
Ambient temperature
Temperature class
T1
T2
T3
T4
T5
T6
Model FCx3xx-F2..., in Class I Div. 2
Ambient temperature
Temperature class
T1
T2
T3
T4
T5
T6
Ambient and process conditions:
T
-20 ... 60 °C (-4 ... 140 °F)
amb
T
-40 ... 60 °C (-40 ... 140 °F) (only for integral mount design devices)
amb, optional
T
-50 ... 200 °C (-58 ... 392 °F)
medium
Protection class
IP 65, IP 67, and NEMA 4X / Type 4X
IMPORTANT (NOTE)
In the case of the remote mount design, the signal cable between the flowmeter sensor and the transmitter must measure at
least 5 m (16.4 ft) in length. Install conduit seals within 18" (45 cm).
42 CI/FCB300/FCH300-EN Rev. F | CoriolisMaster FCB330, FCB350, FCH330, FCH350
≤40 °C (≤104 °F)
200 °C (392 °F)
200 °C (392 °F)
185 °C (365 °F)
125 °C (257 °F)
85 °C (185 °F)
65 °C (149 °F)
≤40 °C (≤104 °F)
200 °C (392 °F)
200 °C (392 °F)
180 °C (356 °F)
115 °C (239 °F)
80 °C (176 °F)
60 °C (140 °F)
≤50 °C (≤122 °F)
200 °C (392 °F)
200 °C (392 °F)
180 °C (356 °F)
120 °C (248 °F)
85 °C (185 °F)
65 °C (149 °F)
≤50 °C (≤122 °F)
200 °C (392 °F)
200 °C (392 °F)
180 °C (356 °F)
115 °C (239 °F)
80 °C (176 °F)
60 °C (140 °F)
≤60 °C (≤140 °F)
200 °C (392 °F)
200 °C (392 °F)
180 °C (356 °F)
120 °C (248 °F)
75 °C (167 °F)
60 °C (140 °F)
≤60 °C (≤140 °F)
180 °C (356 °F)
180 °C (356 °F)
180 °C (356 °F)
115 °C (239 °F)
75 °C (167 °F)
60 °C (140 °F)