• Honeywell recommends that the sensors be bump tested before each day's
use to confirm their ability to respond to gas. Manually verify that the audible,
visual, and vibrator alarms are activated. Calibrate if the readings are not
within the specified limits.
• Portable safety gas detectors are life safety devices. The accuracy of ambient
gas reading(s) is dependent upon factors such as accuracy of the calibration
gas standard used for calibration and frequency of calibration.
• The detector is designed for use only in potentially explosive atmospheres
where oxygen concentrations do not exceed 20.9% (v/v). Oxygen deficient
atmospheres (<10% v/v) may suppress some sensor outputs.
• Extended exposure of the detector to certain concentrations of combustible
gases and air may stress the detector element and seriously affect its
performance. If an alarm occurs due to a high concentration of combustible
gases, a calibration should be performed. If necessary, replace the sensor.
• High concentrations of certain toxic gases, for example, H
LEL sensor. This effect, known as inhibition, is usually temporary but in
extreme circumstances can impair the sensitivity of the LEL sensor after
any gas exposure that causes alarm in the toxic gas sensors.
• The Honeywell BW ™ Ultra is provided with an antistatic coating over the LCD
window to minimize the risk of ignition due to electrostatic discharge. Periodic
inspection of this coating is required to ensure no degradation, delamination,
abrasions or other deformities to this surface.
• Care must be taken to avoid exposure to excessive heat, harsh chemicals
or solvents, sharp edges and abrasive surfaces. Clean the exterior with a
soft, damp cloth.
• The Canadian Standards Association (CSA) requires the LEL sensor to be
bump tested before each day's use with calibration gas containing between
25% and 50% LEL. The instrument must be calibrated if the displayed
LEL value during a bump test fails to fall between 100% and 120% of the
expected value for the gas.