CARESCAPE ONE User Manual DRAFT 26 October 2017
System introduction
LED
On/standby (yellow)
On/standby (green)
On/standby (unlit)

Battery

The CARESCAPE ONE is designed to operate on battery power when used in transport
or whenever AC power is interrupted. A complete battery management system allows
you to obtain maximum battery performance. On-screen capacity gauges indicate
battery charge condition and capacity.
The Lithium-Ion battery is a rechargeable battery containing Lithium-Ion cells. Each
battery contains an integrated electronic fuel gauge and a safety protection circuit.
The processor within the equipment communicates with both the battery and the
charger.

Battery status indicator

LED
Yellow
Yellow, blinking once every 1 to
5 seconds
Yellow, blinking multiple times per
second
Green and yellow, blinking once
every 1 to 5 seconds
Green
Green and yellow, blinking
multiple times per second

Installing batteries

WARNING
42
Description
The yellow on/standby key should be lit when one of the following occurs:
The device is connected to an external power source and the device is
turned off.
The device is booting up or is in standby.
The green on/standby key indicates the device is ready for operation.
The unlit on/standby key indicates there is no power applied to the device.
Description
Solid yellow indicates the battery is being charged.
Slow blinking yellow indicates one of the following:
The device battery is not present.
A battery is present, but device communication with the battery
cannot be established.
Fast blinking yellow indicates the battery charge capacity is less than
60% of the design capacity.
Slow blinking green and yellow indicates the battery time to empty is 20
minutes or less.
Solid green indicates the device battery is in use.
Fast blinking green and yellow indicates the battery time to empty is 5
minutes or less.
EXPLOSION OR FIRE. Using non-recommended batteries
could result in injury/burns to the patients or users. Only use
batteries recommended or manufactured by GE. The warranty
can be voided if non-recommended batteries are used.