5.5.3

Testing UV transparency

Once the pump cycle has been completed, check the vacuum chamber's
UV transparency as follows:
1.
Open the main measuring window in the spectrometer software.
2.
Select "Optic UV transparency test" from the menu item "Analysis".
Fig. 5-8
3.
Place the displayed (fig. 5-9) and freshly prepared recalibration sample on the
spark stand and fix it in place with the hold-down device.
4.
Switch on the excitation source at the rotary knob on the front of the instrument and
press "Start".
Fig. 5-9
Once the measurement is completed, the instrument will indicate the UV transparency
level.
If the message appears in
for operation.
Fig. 5-10
If the message appears in
Hitachi High-Tech Analytical Science.
Testing the UV transparency
Measuring the recalibration sample
green
(UV purge factor min. 80 %), the spectrometer is ready
Argon purge completed successfully
red
(UV purge factor < 80 %), please contact the service of