Supplying Power To The Device - Philips Respironics E30 User Manual

Ventilation solution
Table of Contents

Advertisement

Note
When positioning the device, make sure that the power cable is accessible because removing power is
the only way to turn off the device.
3.3

Supplying power to the device

Complete the following steps to operate the device using AC power:
1. Plug the external alarm module or UPS connector into the power inlet on the E30 ventilator. See Section
3.4 for additional details.
Warning
The external alarm module or Uninterruptible Power Supply (UPS) must be used. If your device
comes with an external alarm module, the device should be plugged into the facility's emergency
backup power outlet (orange outlet), or a UPS should be used to provide battery back-up in the
event of loss of AC mains power.
Note
There is an accessory clip that can be used to secure the power cord and the external alarm module
or UPS to prevent accidental disconnection. Route the cord through the clip and secure the clip to the
enclosure of the device (and if applicable, the external alarm module) using the supplied screw. See
instructions accompanying the accessory for additional information.
2. Plug the socket end of the AC power cord (included) into the power supply (also included).
3. Plug the pronged end of the AC power cord into an electrical outlet that is not controlled by a wall switch.
Philips Respironics E30 User Manual