Power Management (Pm) Pcba; Power Supply; Internal Battery - Philips Respironics V60 Service Manual

Respironics v60 ventilator
Hide thumbs Also See for Respironics V60:
Table of Contents

Advertisement

PM PCBA

Power supply

Internal battery

1049766 Rev A
3.2.1

Power management (PM) PCBA

Battery charging and management circuitry
Internal supply voltages (supply voltages: 3.3 V, 5 V, 12 V, 35 V).
Fan power and tach monitoring.
Backup alarm control circuitry, including power fail detection.
Power switch control circuitry.
System alarm and reset management.
Electrical interfaces between the CPU PCBA, LCD, and the user
interface (UI) PCBA.
Includes an EEPROM for calibration data, board identification
information, and PM PCBA power-on hours.
3.2.2
Power Supply
Provides ventilator and battery recharging power from AC line voltage.
Converts AC line voltage (100 to 240 VAC, 50 to 60 Hz) into 24 VDC
power.
Includes input over-current, output over-voltage, and output current-
limiting protection.
3.2.3
Internal Battery
14.4-V, 11.5-Ah lithium-ion battery has a run-time of at least six
hours under normal conditions.
Provides operating power when AC power is not available.
Provides charge and temperature status to the PM PCBA.
Internal circuitry monitors battery status, provides self-contained
fault control features, and communicates this information to the PM
PCBA.
V60 Ventilator Service Manual
Chapter 3: Theory of Operation
3-7

Hide quick links:

Advertisement

Table of Contents
loading