Philips Respironics E30 User Manual page 41

Ventilation solution
Table of Contents

Advertisement

Alert
Icon
Oximetry: Good
Connection
(icon only)
Flow Sensor Error:
Limited Therapy
SD Card Removed.
Oximetry: Good
Study
(icon only)
Priority
Description
Status
Displays on the therapy
screen when the blower is
on and 3 seconds of good
connection is detected.
Appears at the beginning
of therapy. This screen will
not display again if the
oximetry sensor is removed
and reapplied unless
therapy is stopped and
restarted.
Notification
Flow sensor malfunction.
Notification
Indicates SD card has been
or Alert
removed from therapy
device and not reinserted
before the start of the
current therapy session.
Notification
After exiting therapy, the
therapy device will display
the Good Oximetry pop-
up screen if the oximetry
sensor maintained a good
connection for at least 4
consecutive hours within a
24 hour window with breaks
in therapy no greater than 1
hour each.
Philips Respironics E30 User Manual
User action
No action needed.
Press the control dial to
acknowledge and clear
the message. Disconnect
the device from power.
Immediately remove patient
from ventilator and connect
them to alternate source of
ventilation. Contact Philips
Respironics or the product
support team.
Reinsert SD card, or press the
control dial to clear alert.
Press the control dial to
acknowledge and clear the
message.
Alarms, alerts, and troubleshooting
35