Safety Instructions for Laser Products

The CRxFlex is a Class 1 Laser Product. It uses one laser diode of a 120 mW type,
classification class IIIb, wavelength 640-670 nm. The laser beam's deflection frequency is
120 - 170 Hz. The laser beam divergence is 12 mrad.
Under normal operating conditions – CRxFlex with all covers - there can be no laser radiation
outside the CRxFlex.
The technical concept does not allow the user to remove the top cover. The concept provides
maximum reliability that no image plate jam can occur in the post scan area.
However, the user is allowed to open the front cover, e.g. to solve cassette or image plate
jams at the front side. When opening the front panel, all motor driven system movements will
be stopped (including the laser.
Warning: Risk of injury.
User interventions other than those described in this manual can be hazardous
with regard to laser radiation.

Cleaning of the CRxFlex

Caution: Device damage possible.
Make sure no liquid gets in the CRxFlex.
To clean the exterior of the CRxFlex:
1. Switch off the CRxFlex.
2. Remove the power plug from the socket.
3. Wipe the exterior of the CRxFlex with a clean, soft, damp cloth. Use a mild soap or
detergent if required but never use ammonia-based cleaner.
4. Plug the power plug into the socket.
To clean the interior of the CRxFlex please refer to section 6 of this manual.
CRxFlex Operating Manual V1.1
17