Immunity Test
IEC 60601 Test Level
IEC 61000-4-8
30 A/m
Electromagnetic
compatibility
(EMC). Part 4-8:
Testing and
measurement
techniques. Power
frequency (50/60
Hz) magnetic field
immunity test
NOTE: UT is the AC mains voltage prior to application of the test level.
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it
is used in such an environment.
Immunity Test
Conducted RF IEC 61000-
4-6
Electromagnetic
compatibility (EMC). Part
4-6: Testing and
measurement techniques.
Immunity to conducted
disturbances, induced by
radio-frequency fields.
Radiated RF
IEC 61000-4-3
Electromagnetic
compatibility (EMC) - Part
4-3: Testing and
measurement techniques
- Radiated, radio-
frequency,
electromagnetic field
immunity test
Compliance Level
30 A/m
IEC 60601 Test Level Compliance Level
3 Vrms
3 Vrms
150 kHz to 80 MHz
150 kHz to 80 MHz
6 Vrms
6 Vrms
Amateur Radio &
Amateur Radio & ISM
ISM Bands between
Bands between 150 kHz
150 kHz and 80 MHz
and 80 MHz
10 V/m
10 V/m
80 MHz to 2.7 GHz
Up to 28V/m in
Up to 28V/m in
telecommunication
telecommunication bands
bands as specified in
as specified in clause 8.10
clause 8.10 of IEC
of IEC 60601-1-2
60601-1-2
EMC Information
Instructions for Use | Technical Specifications
Electromagnetic Environment - Guidance
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical hospital
or home environment.
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any part of
the device, including cables, than the
recommended 30 cm separation distance.