SIMV-PC: Synchronous Intermittent Mandatory Ventilation- Pressure Control
D
:
ESCRIPTION
SIMV-PC mode is a pressure control mode that provides a mixture of mandatory and spontaneous breaths. SIMV-PC mode guarantees one
mandatory breath in each cycle. Spontaneous breaths can be delivered with pressure support. The breath rate determines the length of the cycle.
The first phase of the cycle is reserved for synchronizing a mandatory breath with patient effort. If the patient triggers a breath during this phase of
the cycle, the ventilator delivers a synchronized mandatory breath. If a patient does not trigger a breath during the mandatory phase of the cycle,
then the ventilator delivers a mandatory breath. Breaths triggered by the patient after the mandatory breath in the cycle are spontaneous breaths.
This process is repeated at the start of every cycle.
S
:
ETTINGS
Setting Name
Description
Pressure Control
Defines the applied pressure for all breaths
Pressure Support
Target pressure that the device delivers during the inspiratory phase of a spontaneous breath
PEEP
Positive end expiratory pressure
Positive pressure maintained in the patient circuit during exhalation: must be less than or equal to the
pressure setting.
Inspiratory Time
For a mandatory breath, length of the inspiratory phase
Rise Time
Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure
setting when the breath is triggered.
Breath Rate
Minimum rate of mandatory breaths per minute
Trigger Type
Auto-Trak
Sensitive Auto-Trak (passive circuits only)
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits)
Flow Trigger
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patient's
Sensitivity
inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting.
Flow Cycle
This control is available when the trigger type is Flow Trigger
Sensitivity
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the
device cycles to expiration.
S
:
ETTABLE ALARMS
Circuit disconnect
High tidal volume
Low tidal volume
High minute ventilation
Low minute ventilation
High respiratory rate
Low respiratory rate
I
LLUSTRATION
Instructions for Use | Therapy Modes and Controls