Philips Trilogy Evo Universal Instructions For Use Manual page 94

Table of Contents

Advertisement

Guidance and Manufacturer's Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it
is used in such an environment.
Immunity Test
IEC 60601 Test Level
IEC 61000-4-2
±8 kV contact
Electromagnetic
±2 kV, ±4 kV, ±8 kV, and
compatibility
±15 kV air
(EMC). Part 4-2:
Testing and
measurement
techniques.
Electrostatic
discharge
immunity test
IEC 61000-4-4
±2 kV for power supply lines
Electromagnetic
±1 kV for input-output lines
compatibility
(EMC). Part 4-4:
Testing and
measurement
techniques.
Electrical fast
transient/burst
immunity test
IEC 61000-4-5
±1 kV line to ground
Electromagnetic
±2 kV line to ground
compatibility
(EMC) - Part 4-5
Testing and
measurement
techniques. Surge
immunity test
IEC 61000-4-11
O% U
T
Electromagnetic
increments
compatibility
0% U
T
(EMC). Part 4-11:
70% U
Testing and
(30 cycles if US)
measurement
0% U
T
techniques.
Voltage dips, short
interruptions and
voltage variations
immunity tests
Compliance Level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
and
±15 kV air
±2 kV for supply
mains
±1 kV for
input/output lines
±1 kV line to line
N/A - this Class II
device does not
connect to earth
ground
0.5 cycle at 45 degree
O% U
degree increments
1 cycle
0% U
25 cycles
70% U
T
(30 cycles if US)
5 sec
0% U
EMC Information
Instructions for Use | Technical Specifications
Electromagnetic Environment - Guidance
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical home
or hospital environment.
Mains power quality should be that of a typical home
or hospital environment.
0.5 cycle at 45
Mains power quality should be that of a typical home
T
or hospital environment. If the user of the device
1 cycle
requires continued operation during power mains
T
25 cycles
interruptions, it is recommended that the device be
T
powered from an uninterruptible power supply or a
5 sec
battery.
T
94