  • Page 1 DreamStation Go Heated Humidi er User manual Instructions for use...
  • Page 2: Table Of Contents

    Physical ....................................17 Service life ....................................17 Standards compliance ..............................17 IEC 60601-1 classification ..............................18 Electrical (when the heated humidifier is used with the DreamStation Go therapy device) ....18 Maximum recommended pressure ..........................18 Water capacity ..................................18 Heater plate ..................................18 Inhalation ressure drop of humidifier ..........................
  • Page 3: Safety Information

    Safety information Intended use The DreamStation Go Heated Humidifier is an accessory for the Philips Respironics DreamStation Go therapy devices to provide moisture to the patient circuit. It is intended for use in spontaneously breathing patients weighing over 30 kg (66 lbs), in the home or hospital/institutional environment, who use mask-applied positive pressure ventilation therapy.
  • Page 4 Use a mild liquid dishwashing detergent only for either hand washing or when using a dishwasher. • Only the hospital and institution cleaning procedures listed in this manual are recommended by Philips Respironics. Use of other cleaning and disinfecting processes, not specified by Philips Respironics, may affect the performance of the product. Note The DreamStation Go Heated Humidifier is specified for use with Philips Respironics DreamStation Go therapy devices only.
  • Page 5: System Overview

    Murrysville, PA 15668 System overview The DreamStation Go Heated Humidifier attaches to the therapy device and provides an air outlet port to connect a breathing circuit. The breathing circuit is composed of patient tubing, a mask, and in some instances a separate exhalation device.
  • Page 6 System features and contents The figure below illustrates many of the device features and contents. In addition to the features shown below, your humidifier system may include a humidifier user manual. Heated humidifier Water tank lid Air outlet port Manifold Water tank lid tab Baffle CPAP air inlet...
  • Page 7: Connecting And Setting Up The Humidifier

    Connecting and setting up the humidifier Connecting to the therapy device Remove the humidifier beauty panel from the therapy device by pulling from either side of the panel. Pull here Caution 2. Place the therapy device and heated humidifier (with an empty water tank) on a firm, flat surface.
  • Page 8 Setting humidification mode and target time To access the humidifier settings in the Therapy Off menu, navigate to the My Comfort menu. From here you can view or adjust the following humidifier settings. Icon Text Description Humidification This displays the humidification mode being used. You can choose between Fixed or Adaptive humidification.
  • Page 9: Humidifier Pop-Up Messages

    Humidifier pop-up messages Device pop-ups are messages that show up on the user interface screen. Condition Icon Description Possible Action Cause Humidification Humidifier error (only Humidifier heater Turn off the device and Error when humidifier is plate error disconnect it from power. present).
  • Page 10: Daily Use

    — Do not put any chemicals or additives into the water. Possible airway irritation or damage to the water tank may result. — Philips Respironics recommends using room temperature distilled water, if available, to prevent mineral build up on the humidifier tank base. In a clinical environment, distilled water must be used.
  • Page 11 4. Remove the tank lid by pulling up on the tank lid tab while holding the tank base. Pour out any remaining water and rinse the tank base. 5. Place the tank base on a firm, flat surface. Fill the tank base with water no higher than the maximum fill line Note...
  • Page 12 Disconnecting the humidifier Caution To avoid spilling, do not disconnect the humidifier from the therapy device with water in the tank. Remove the water tank from the humidifier before disconnecting the therapy device. Disconnect power to the therapy device and tubing.
  • Page 13: Cleaning

    A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan. Note Refer to the DreamStation Go therapy device user manual for tube cleaning instructions. Cleaning the water tank base Hand washing can be performed daily. Dishwashing can be performed weekly.
  • Page 14 2. Remove the baffle (2) by pressing in on the side tabs and pulling up on the baffle. 3. Wash the lid and components in a solution of warm water and a mild dishwashing detergent. Rinse with clean water. 4. Allow components to air dry. 5.
  • Page 15: Maintenance Instructions

    Maintenance instructions Warning Allow the humidifier heater plate and water to cool down for approximately 15 minutes before removing the water tank. A burn may result from: touching the heater plate, coming in contact with the heated water, or touching the tank pan. Maintaining the water tank base If you use tap water in your humidifier water tank, follow the steps below to maintain the tank base.
  • Page 16: Troubleshooting

    The therapy device The humidifier has an Contact your supplier or Philips Respironics. If the problem cannot is operating but the airflow obstruction. be resolved, return the humidifier to your supplier to determine the humidifier’s airflow is problem.
  • Page 17 Confirm that the humidifier electrical contacts are not obstructed or damaged. Then reconnect the humidifier and reconnect the device’s power supply. Turn the blower on; if the settings are still not visible, contact your supplier or Philips Respironics for assistance. The water in the water Water tank was...
  • Page 18: Additional Notes

    Contact your supplier for additional information. Service The humidifier does not require routine servicing. If any part of the humidifier is worn or damaged, contact Philips Respironics or your supplier. See the Troubleshooting section in this manual for additional information.
  • Page 19: Dreamstation Go Heated Humidifier Specifications

    Service life The expected service life of the DreamStation Go heated humidifier is five (5) years. The expected service life of the DreamStation Go heated humidifier water tank is one (1) year. Standards compliance This device is designed to conform to the following standards: •...
  • Page 20: Iec 60601-1 Classification

    Explanation: Protected against vertically falling water drops when enclosure tilted up to 15°. Mode of Operation Continuous Electrical (when the heated humidifier is used with the AC Power Consumption: 100–240 VAC, 50/60 Hz, 2.0- DreamStation Go therapy device) 1.0 A Maximum recommended pressure 20 cm H Water capacity 240 ml (8 oz.) at recommended water level...
  • Page 21: Declared Dual-Number Noise Emissions Values

    Declared dual-number noise emissions values (in accordance with ISO 4871) Tube Size Sound Pressure Uncertainty (K) Sound Power Uncertainty Level (L) Level 12 (mm) tubing type 33.9 dB(A) 2.0 db(A) 41.9 dB(A) 2.0 dB(A) 15 (mm) tubing type 33.3 dB(A) 2.0 db(A) 41.3 dB(A) 2.0 dB(A)
  • Page 23: Limited Warranty

    Limited Warranty Respironics, Inc., a Philips company, warrants that the system will be free from defects of workmanship and materials and will perform in accordance with the product specifications from the date of sale by Respironics, Inc. for a period of two (2) years.