c) On the front shell, ensure the LCD gasket is aligned
with the pins. The frame should be facing up.
7. Attach the front shell:
a) Place the PCB into the front shell and insert the two
screws (torque 3-4 in-lbs) to fasten both pieces.
b) Replace the back shell and install the four screws
back into the battery compartment (torque 4-5 in-lbs).
c) Reinstall the 2 screws at the top of the instrument
(torque 4-5 in-lbs).
8. Attach the battery:
a) To place the battery pack correctly, first engage
hooks on the bottom of the battery pack.
BW Ultra User Manual
b) Push the battery pack into place and tighten the
battery retention screw (torque 4-5 in-lbs).
c) Turn on the instrument and allow for the sensors to
stabilize.