Wiring The Remote Start (Binary Input) Connection; Figure 6 Wiring Diagram For The 4-20 Ma Connection - GE Sievers 900 Series Operation And Maintenance Manual

On-line total organic carbon analyzers
Table of Contents

Advertisement

Internal to 900 On-Line Analyzer
4-20 mA Output
Isolated self power
Voltage compliant to 20 V
Pin 12 (-)
Pin 11 (+)
TB3

Wiring the Remote Start (Binary Input) Connection

The 900 On-Line Analyzer remote start (binary input) connection can be wired in a variety of ways
and is intended to execute a start/stop command to the Analyzer from a remote location, such as a
Process Logic Control System or Center (PLC). Wiring examples are shown in Figure 7 and Figure 8.
Note that binary input must be activated in the Analyzer firmware later. It is important to confirm
the correct signal strength before setting binary input to On, to avoid unexpectedly starting or
stopping analysis. For more information on activating binary input, see "Activating Binary Input" on
page 110.
GE Analytical Instruments ©2009
Figure 6: Wiring Diagram for the 4-20 mA Connection
Max. Load 600 ohms
No external power required
(-)
(+)
or
Recorder
71 of 236
Rload
DLM 90488-01 Rev. A