Satellite XT R Version
Used to define the trigger condition for Alarm 1
activation. Default setting is Greater Than. Use the
<set> key to enter. The instrument will show the actual
o f
4
setting.
Defines that an Alarm 1 condition will be indicated when
the actual gas concentration exceeds the preset level
for Alarm 1.
1
o f
2
Press the cursor keys <up> or <down> to change.
Press the <set> key to confirm or <esc> to exit..
Less Than
Defines that an Alarm 1 condition will be indicated when
the actual gas concentration falls below the preset level
for Alarm 1.
2
o f
2
Press the cursor keys <up> or <down> to change.
Press the <set> key to confirm or <esc> to exit..
Used to define the Alarm 1 activation to be latching or
non-latching. Default setting is enabled. Use the <set>
key to enter. The instrument will show the actual setting.
o f
4
When enabled, alarm reset must be done manually.
Non-latching alarms will be reset automatically, upon
correction of the alarm condition.
Defines the Alarm 1 activation to be non-latching. Press
the cursor keys <up> or <down> to change. Press the
<set> key to confirm or <esc> to exit.
1
o f
2
Defines the Alarm 1 activation to be latching.
Press the cursor keys <up> or <down> to change.
Press the <set> key to confirm or <esc> to exit.
2
o f
2
Configuration - Section 6
Alarm 1 Trigger
Greater Than
Alarm 1 Latch
Disabled
Enabled