The Nurse Call Relay - Philips V24CT User's Reference Manual

Cms patient monitoring system volume 1
Table of Contents

Advertisement

The Nurse
Note—
Call Relay
The nurse call relay is a paging system which warns the nurse if an alarm
occurs on a monitor. It becomes active in one of four situations
depending on the way it has been configured:
The nurse call relay is not active when the monitor's alarm capability is
switched off.
Note—
When an alarm has been silenced using the
audible 2 or 3 minute alarm reminder is configured to be either On or Re-
alarm, the nurse call relay signals will behave as follows:
The nurse call relay is not available for the V24CT and V26CT.
Red only alarms
Red and Yellow alarms
Red and INOP alarms
Red and Yellow INOP alarms.
The delay time to the nurse call relay output is max. 1 sec.
Alarm
Reminder
Reminder
Time
On
2 or 3 mins
Re-alarm
2 or 3 mins
Silence/Reset
Nurse Call Relay Signal Behavior
The nurse call relay signals are repeated
for a duration of 6 seconds every 2 or 3
minutes unless the alarm condition is
eliminated
The nurse call relay signals are activated
after 2 or 3 minutes. The alarm tone and
relay signals run continuously until they
are interrupted by pressing the
key.
Silence/Reset
Alarm Setup
key, and the
Alarm Functions 5-15