Angulation Brake Magnet - Siemens SIREMOBIL Iso-C Troubleshooting Manual

Table of Contents

Advertisement

138

Angulation brake magnet

Fig. 54:
CAUTION
Remove the cover of the horizontal carriage.
Remove the back cover with the handle for the horizontal lift movement after removing
the 4 countersunk screws. The angulation brake magnet is now visible.
Remove the cover plate over the C-arm cable guide.
Remove the three attachment screws and remove the brake magnet.
NOTE
Remove the plug connection for the brake magnet and insert the new brake magnet.
Place the new brake magnet back on the shaft and attach it with the three screws.
Reattach the cable with cable ties.
Reattach the cover over the C-arm cable guide.
Reinstall the cover. Make sure the protective conductor is connected properly.
Perform a function test of the brake. See
SIREMOBIL Iso-C
Risk of burns!
See
(Safety information / p.
When the angulation brake is released, the brake mag-
net heats up. Be sure to switch off the power supply
and let the brake magnet cool off before touching it.
Remove both parts of the brake magnet!
SPR2-230.840.01.01.02
08.08
Brakes / Movements
11).
(Brake force / brake torque / p.
Page 138 of 148
CS PS SP
7.1
135).
© Siemens, 2008
For internal use only