Gantry Software Updates
Procedure
14. Repeat steps 1 (page 7-4) to 12 (page 7-5)
of this procedure to update the micro con-
troller firmware of the EDC module on the
other detector.
15. When both EDC modules have been up-
dated, remount the detector top cover fol-
lowing instructions provided in Chapter 6.
Repair Procedures.
Upgrading Logic Firmware
Procedure
To download logic firmware to the EDC
board, you use the terminal emulator program
ttermpro. Proceed as follows:
1. In the Windows Explorer, select the folder
EDC CardioMD programming for
Module rev. 9CSY0900-
on the CD-ROM.
A01\TTERMPRO
2. In this folder, double-click the icon
to start the ttermpro pro-
ttermpro.exe
gram.
3. Press [Enter] to check if the terminal
prompt
is present. If the prompt does
->
not appear, check the serial port setup:
Baudrate: 57600
Data: 8
Parity: none
Stop: 1
Flow control: hardware (or none)
Transmit delay: 0 and 1 communication.
4. At the terminal prompt, type the following
commands:
-> dspreset
-> ispsvf
5. In the File menu, select Send File.
6. Select the file
LSI.svf
CardioMD programming for Module
rev. 9CSY0900-A01
and click Open.
A pop-up appears counting the number of
bytes transferred. When the firmware is suc-
cessfully loaded, the pop-up closes automati-
cally. At this point, the EDC module no longer
7-6 CardioMD Service Manual
Details
Details
in the folder
EDC
on the CD-ROM,
1MAN0158-H11 04-2005 CSIP Level 1