Selecting the ECG filter
ECG
1. Press the
key.
2. Select ECG Setup - Filter, and select the appropriate option:
STfilt: Filters high-frequency artifacts but catches slow ST
changes.
Monit: Filters high-frequency artifacts and slow ST changes.
Diagn: Catches high-frequency changes and slow ST
changes.
Changing the HR source
ECG
1. Press the
key.
2. Select ECG Setup - HR Source, and select the appropriate option.
AUTO, which selects the first available HR source
ECG
Art
ABP
UAC
Pleth
NOTE: Art, ABP, and UAC display when a related label is selected.
Selecting other HR parameters
You can select what is displayed with heart rate:
ECG
1. Press the
key.
2. Select ECG Setup - Display with HR, and select the appropriate
option.
PR
PVC
None.
Displaying the ECG grid
ECG
1. Press the
key
2. Select ECG Setup - Select Grid, and select the appropriate
option:
ON: To view ECG wavefroms over gridlines
OFF: To view ECG wavefroms without gridlines
Adjusting the beat sound volume
ECG
1. Press the
key.
2. Select ECG Setup - Beat Sound Volume.
3. Adjust the volume from 0 to 10. If you select 0, there is no audible
sound.
Setting the ECG alarms
ECG
1. Press the
key.
2. Select ECG Alarms.
To set up the limits, select Adjust Limits.
To OFF/ON the HR limits. select HR Alarm.
To OFF/ON the PVC limits, select PVC Alarm.
To adjust ST limits, select ST Alarms.
3. You can also select Alarms Setup to adjust ECG or related
alarms.