Audible alarms off behavior
When audible alarms are turned off:
All audible alarms are turned off except for specific alarms
configured to break through the audio off setting.
The audio off bell icon displays in the upper left corner of the
display screen.
The visual alarm signals still display.
Turning audible alarms on/off
Alarms Setup
1. Press the
option is not selectable, see
2. Select an alarm group. Choices are:
Silence Apnea: Turns off audible alarms for apnea,
respiration rate, EtCO2 and FiCO2 limit alarms, except the
related technical alarms.
Silence ECG: Turns off audible alarms for all HR source (ECG,
Art and Pleth) limit and arrhythmia alarms, except the related
technical alarms.
Silence Apn&ECG: Turns off audible alarms for apnea,
respiration rate, EtCO2, FiCO2, all HR source limit and
arrhythmia alarms, except the related technical alarms.
Silence ALL: Turns off all audible alarms (both of
physiological and technical), except specifically defined as
breakthrough alarms.
3. To turn on all audible alarms again, select Activate Alarms
NOTE: If alarms are turned off for any of the defined alarm groups
and an alarm occurs within the alarm group, a beep tone will sound
every 2 minutes as a reminder that alarms are turned off.
key and select Audio ON/OFF. If this
"Changing alarm
options."
43