Additionally, on the CARESCAPE network, the monitor supports
remote parameter and waveform configurations from the central
station.
NOTES:
•
The Entropy parameter is not supported by CIC pro and
CARESCAPE Central station.
•
On the iCentral, the care area settings can be setup to "Non OR"
or "OR" in the DRI service menu. Please consult the authorized
service personnel.
Non OR:
−
Displays up to 72 hours of trend data.
NOTE: Anesthesia Gas trends are not supported
−
Does not display Entropy alarm limits
OR:
−
Displays up to 24 hours of trend data
24
Connection to Network
Ensure that the appropriate network infrastructure is in place prior
to the installation of the patient monitor.
NOTE: The network installation and configuration are performed by
authorized service personnel only.
MC and S/5 Network
Use the CAT-5 network cable to connect the monitor to the network.
1. Make sure that the power is switched off.
2. Connect one RJ-45 connector to the network port at the back of
the monitor.
3. Connect the other RJ-45 connector to the corresponding port on
the wall.
4. Turn on the power. Confirm that the network symbol and
'Network: xxx' message are displayed.