Admit Menu - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Admit Menu

2000966-542D
Configuration: Service Mode settings
The Admit Menu selection determines the function of the patient
monitor. Before configuring the Admit Menu, you must know if the
patient monitor will be used for standard adult, neonatal, or operating
room monitoring, and if the patient monitor will be moved from room to
room. All combinations are explained below.
STANDARD—configures the patient monitor to stay in one room for
stationary monitoring only. Patient monitors not connected to the
CARESCAPE Network (Ethernet connection) must use STANDARD
configuration only.
ROVER—configures the patient monitor to move from room to room
for stationary monitoring only.
COMBO—configures the patient monitor to stay in one room for
both stationary and ambulatory (telemetry) monitoring. This patient
monitor displays all Tram module data combined with ECG data for
ambulatory patients.
NOTE
Users should be aware of a possible time discrepancy between
the waveforms from the Telemetry device and the waveforms
hard-wired to the Dash patient monitor. Users should not
consider these waveforms to be synchronous. If absolute
synchronicity is desired, Combo mode should be discontinued and
the ECG waveforms should be acquired via the hard-wired
bedside patient monitor.
ROVER COMBO—configures the patient monitor to move from
room to room for both stationary and ambulatory (telemetry)
monitoring.
1. Enter service mode (See
select MENU SETUP > ADMIT MENU.
2. Use the Trim Knob control to select the function of the patient
monitor.
3. Press Trim Knob control to exit.
Dash 3000/4000/5000
"Service Mode settings" on page
513A
4-11) and
4-13