Remove or replace handle assembly

7-10
Field replaceable units: Remove or replace handle assembly
1. Remove the two screws holding the handle to the patient monitor.
Take care not to strain the cables if your patient monitor has the
alarm light option or the wireless option.
2. If your patient monitor does not have the alarm light option or
wireless option, replace the defective handle assembly or continue to
"Remove or replace display assembly"
parts.
3. If your patient monitor has the alarm light option, replace the
defective handle assembly, or continue to remove the display
assembly to route the alarm light cable through the Dash housing (if
it is not already routed) or to replace other parts. Refer to
replace display assembly"
NOTE
For Dash 3000 only: If the replacement handle assembly has an
alarm light, the DAS assembly must be removed to connect the
cable to J1 on the writer board. Refer to
on page 7-37 for instructions on removing and reinstalling the
DAS assembly.
Dash 3000/4000/5000
on page 7-18 to replace other
on page 7-18.
"Replace DAS assembly"
alarm light cable
537A
"Remove or
967A
2000966-542D