Patient data storage
Time and date
Calibration data
Error log

Optional thermal printer

2000966-542D
Equipment overview: Theory of operation
Static RAM backed up by a lithium battery soldered onto the processor/
power management PCB stores patient data.
A patient monitor on the network maintains time and date for at least
five years. Time is accurate to within one second in ten hours. The clock
is synchronized with a time master on the optional CARESCAPE
Network.
Calibration factors for NBP and CO
on the DAS board.
Calibration factors for the analog output signals are stored in non-
volatile memory on the processor/power management PCB.
Input and output errors are retained in static RAM backed up by a
lithium battery soldered onto the processor/power management PCB.
Contents are retained for at least five years, provided that you exercise
caution when you handle the board to prevent inadvertently discharging
the battery (e.g. when you ship the board for problem diagnosis).
The patient monitor uses the same 50-mm thermal printer module that
is used in the PRN 50 stand-alone printer. It prints up to four waveforms
at chart speeds ranging from 0.1 to 50 mm/s. The printer software is
loaded independently from the patient monitor's software.
In the patient monitor, the printer module limits its current consumption
to stay within its allocated system power budget.
Dash 3000/4000/5000
are stored in non-volatile memory
2
2-43