Test Procedure "Blood Pressure (Nibp) - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

t
"B
est proCedure
lood pressure
Unless regulatory requirements of the country define a different test procedure or test interval, it is recommended to perform the
test "Blood pressure (NIBP)" at least every 2 years
A calibrated NIBP measure instrument set is needed to perfom the following test procedures
Please refer to the operator manual of the used device
t
est preparatIon
1. Power eBike III on and enter service menu / NIBP test (-see „Calibration of Blood pressure unit" on page 75).
2. Connect the calibration equipment with the eBike III cuff connector.
l
-
eak
tIgHtness test
1. Use the connected test equipment to apply the target pressure (see table below) to the eBike III cuff connector, wait
until the pressure in the system has stabilized (approx. 5 sec).
2. After 1 min check the actual pressure and note the pressure drop in the table
Step
Target Pressure
#1
50 mmHg
#2
300 mmHg
a
CCuraCy test
1. Use the connected test equipment to apply the target pressure (see table below) to the eBike III cuff connector, wait
until the pressure in the system has stabilized (approx. 5 sec).
2. Check the actual pressure displayed on the eBike III display and note the value in the table.
Step
Target pressure
#1
0
#2
50
#3
100
#4
150
#5
200
#6
250
#7
300
2018111-336 Rev. B
(nIBp)"
Note
Note
Actual pressure drop
mmHg / min
mmHg / min
Displayed pressure (eBike III)
eBike III basic / comfort
Pressure drop max.
Pass / Fail
< 6 mmHg / min
< 6 mmHg / min
Deviation max.
Pass / Fail
± 3 mmHg
± 3 mmHg
± 3 mmHg
± 3 mmHg
± 3 mmHg
± 3 mmHg
± 3 mmHg
- 17 -