(6) Measurement start input = CCD trigger, start sampling = parallel, object type change =
parallel, result output = computer link/parallel
When the start sampling input (X0) is turned ON, the CCD trigger is enabled. (Sampling starts)
Setting order 1
1MEAS INP I/F
PARALLEL+SERIAL
TRIG CCD START
2 START CCD SAMPL
3 OUT I/F(PARAL.)
Note 1: When the settings listed in section "16-3 CCD trigger" have not been made, a "CCD TRIG
NOT SET. (error 34)" will occur.
Note 2: Start sampling input (X0)
1. Sampling will be performed while this input terminal is ON. When it is turned OFF, the
sampling will stop. During sampling, O will flash in the upper right corner of the operation
screen.
2. After the measurement is terminated, the sampling will be restarted when the X0 terminal
is changed from OFF to ON.
Time chart
Measurement
start input
(X0)
Start CCD sampling by turning ON
measurement start input (X0).
CCD trigger
cycle
Type selection
input
Specified object type number
See page
16-5.
Computer link
CCD trigger
Black
White
window
(not highlighted)
Trigger
READY output
(While measuring or setting = L)
Terminate
measurement
(C119)
P
P
P
P
Halt PC operation
PC calculation
C
C
C
C
during measurement
condition
When the object type is changed,
all Y output and auxiliary relays
turn OFF.
Parallel output
(Y0 to Y7)
When a Sharp PC is used, a write enable command (EWR) is transmitted from the IV-S30J to the
PC in the following cases.
- When the power is applied to the IV-S30J.
- When a write mode nonconformity error (code 10
is transmitted (when the power is disconnected from the PC).
- When the output method is changed from the "OUT I/F (PARAL.)" to the "COMPUTER LINK."
2 (
3)
AUTO(EDGE)
AUTO(LEVEL)
PARALLEL
SERIAL
NO
PC-LINK
SERIAL
Result
output
READY signal is turned ON
at the end of
the result output.
Measurement result is valid at the
end of a measurement
P
P
P
P
P
P
P
P
P
P
C
C
C
C
C
C
C
C
C
C
Parallel output is valid for PC control after
the measurement result is determined.
Result output
16-11
Setting the Input/Output Conditions
Configuration example
Camera 1 (image capturing + CCD trigger)
Camera 2 (image capturing)
Data
Remote keypad
(Computer link)
Controller
Power supply
(24 VDC)
Programmable
controller
Object type change
(parallel I/F)
-See page 16-5.
Result
output
Black
Trigger
P
P
Halt PC operation
P
P
P
P
P
P
C
C
during measurement
C
C
C
C
C
C
Result output
) occurs after a result write command (WRG)
(H)
Monitor
Start sampling
input X0
Halt CCD
sampling by
turning OFF
measurement
start input
(X0).
P
P
P
P
C
C
C
C
16