Chapter 3 Features and Functions
Measurement Configuration
Continuity Threshold Resistance
When measuring continuity, the multimeter emits a continuous tone if
the measured resistance is less than the threshold resistance. You can
set the threshold to any value between 1
The threshold resistance is adjustable only from the front panel.
The threshold resistance is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
The factory setting for the threshold resistance is 10 .
After enabling the continuity function, you can select a different
threshold resistance by pressing
2: CONTINUITY
See also "To Test Continuity," on page 19.
52
and 1000 .
(Menu Recall).
Shift >
(MEAS MENU)
0010
OHM