HP 34401A User Manual page 120

Hide thumbs Also See for 34401A:
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
range
resolution
CONFigure:FREQuency {<
>|MIN|MAX|DEF},{<
>|MIN|MAX|DEF}
Preset and configure a frequency measurement with the specified range
and resolution. This command does not initiate the measurement.
For frequency measurements, the multimeter uses one "range" for all
inputs between 3 Hz and 300 kHz. With no input signal applied,
frequency measurements return "0".
CONFigure:PERiod {<
range
>|MIN|MAX|DEF},{<
resolution
>|MIN|MAX|DEF}
Preset and configure a period measurement with the specified range
and resolution. This command does not initiate the measurement.
For period measurements, the multimeter uses one "range" for all
inputs between 0.33 seconds and 3.3 sec. With no input signal applied,
period measurements return "0".
CONFigure:CONTinuity
Preset and configure the multimeter for continuity measurements.
This command does not initiate the measurement. The range and
resolution are fixed for continuity tests (1 k range and 5
1 2
digits).
CONFigure:DIODe
Preset and configure the multimeter for diode measurements.
This command does not initiate the measurement. The range and
resolution are fixed for diode tests (1 Vdc range with 1 mA current
source output and 5
1 2
digits).
CONFigure?
Query the multimeter's present configuration and return a quoted string.
120