Power-Up/Self Test Messages - HP 53131A Operating Manual

225 mhz universal counter
Hide thumbs Also See for 53131A:
Table of Contents

Advertisement

NOTE
Chapter 2 Operating Your Universal Counter
Front Panel Display Messages

Power-Up/Self Test Messages

The following sequence occurs when you turn on the Counter:
1. All of the vacuum-fluorescent display segments and the indicators
(or LEDs) are illuminated for the power-up test.
2. If all tests pass: SELFTST: PASS will be displayed (See NOTE below).
3. HP-IB AT X will be momentarily displayed (where X represents the
HP-IB address number), then the Counter will start taking
measurements. (Note: Earlier versions of the Counter do not display
the HP-IB address.)
If there is a failure, each test that fails will display a failure message, and
SELFTST: FAIL will be displayed. You must press any key in order for the
Counter to go on.
Possible individual failure messages are:
FAIL: DISP
FAIL:CPU
FAIL: ROM
FAIL:RAM
FAIL: EEPROM
FAIL: HP-IB
FAIL: QSPI
FAIL: FPGA
FAIL: FR END
FAIL: MEAS
FAIL: INTERP
In addition, the following power-up message may be displayed:
UNCALIBRATED
This means that at least one type of calibration has NOT been performed.
2-74
display test
processor test
ROM test
RAM test
EEPROM test
HP-IB hardware test
serial hardware loops test
count circuit test
front end test
measurement test
interpolator test
Operating Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53132a

Table of Contents