To Secure Against Calibration - HP 53131A Operating Manual

225 mhz universal counter
Hide thumbs Also See for 53131A:
Table of Contents

Advertisement

Chapter 2 Operating Your Universal Counter
Using the Calibration Menu
Out of HP 8130A, into HP 59992A J06 Calibrator Signal:
PERIOD: 100 ns
WIDTH: 50 ns
High: 0.50 V
Input Mode: Normal
Timebase: Any external timebase you provide is ignored during
calibration.
Procedure: From the front-panel calibration menu, one keypress
invokes the calibration. You are prompted to press buttons on the
HP 59992A J06 Calibrator for each part of a four-part calibration.
Note that CAL: TI FINE? requires the completion of four steps in
order.

To Secure Against Calibration

1
Press and hold Scale & Offset key, then cycle POWER key.
CAL UNSECURE is displayed. If CAL SECURE is displayed, then the
Counter is already secured against calibration and the remaining steps
are not applicable.
2
Press Scale & Offset key until CODE: is displayed.
3
Press the appropriate arrow keys to enter the security code, then press
Enter key.
CAL SECURE is displayed.
Operating Guide
Low: 0.50 V
2-71