Before You Begin
Figure 1-1: ANCB Board
Jumper Settings
J11 Port 2
For ANCB Ju mp ers on
Botto m 3 po sitions On ly
NAU Installations
Instructions
14 • Installation
This section provides instructions for connecting the Maxum to a
DataNET network. All DataNET connections are made to the ANCB. The
DataNET connections can be either copper or fiber.
The ingress wiring method (conduit or cable glands) and which cable
type you use is dependent upon the site requirements. Review the Site
Wiring Requirements page 8 to ensure that you are using the correct
cable and wiring method.
Refer to ANCB Description page 46.
J 10 P ort 1
For ANCB: Ju mpers on
Botto m 3 Position s O nly
J12
J4
10B ase2
10B aseT
J5
Default on Pins 1 &2
For rack mount units the SYSCON connections are visible at the rear of
the NAU. If you have a wall mounted unit you must first remove the front
or side safety box access covers to see the SYSCON connections.
Step
1.
Turn off AC Power (Mains) to the unit.
2.
Route incoming the DataNET copper cables to the ANCB
in accordance with national or local electrical codes and
regulations.
Config Port
RS232
57.6 KB
O peration LED S
J2 Ju mper Co nn ection s
• • 150 Ohm
• • 120 Ohm
Port 2 (DN)
• • 100 Ohm
• • 150 Oh m
• • 120 Ohm
Port 1(DN)
• • 100 Ohm
Default S etting on150 o hm
Position s
Action
Board
RESET
2000592-001