1107-0641-000
Performance
Specifications
Concentration range
1 to 18%
Scale
1% from 1 to 10%
2% from 10 to 18%
Calibrated flow range
at21+2°C
0.2 to 10 liters/minute
Accuracy at 5 liters/
min Os
+ 0.5% v/v of delivered agent or + 15%
dial setting (whichever is greater)
Liquid capacity
390 ml nominal volume (indicated on level
display)
240 mi refill capacity indicator
nominal 60 ml reserve (not indicated on
level display)
Flow resistance
Less than 50 cm H,O at 5 liters/minute of
Os at21 °C
Less than 250 cm H50 at 15 liters/minute
of Os at 21 °C
Battery
Duracell MN 1604 or VARTA Energy 2000
Operating
temperature range
18 °C to 30 °C (64 °F to 86°F)
Operating humidity
range
30% to 75% relative humidity
(non-condensing)
Storage temperature
range
-40 °C to 60 °C (-40 °F to 140°F)
Storage relative
humidity range
0% to 95% relative humidity at 40 °C
(non-condensing)
Storage atmospheric
pressure
70.0 to 106.0 kPa