Operating Instructions
Preoperative checkout
Step 1
Test the vaporizer for leaks.
e
The Tec 6 Plus Vaporizer
can be positive pressure or
negative pressure leak
tested.
O
Leak test
+
Perform the leak test as
described in the relevant
anesthesia system's
operation and
maintenance manual
initially with the dial at O
(©). Repeat the test with
the vaporizer dial turned to
the 1% setting (2) and
then turn the dial to the O
setting (9).
1%
Leak test
Figure 5-6
Testing for leaks
1107-0641-000
1/23/01
5-5