Frequency Counter - Accuracy - HP 3457A Service Manual

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

Performance Tests
4-38. Frequency Counter - Accuracy Test
Frequency Counter Performance Tests Cont'd
3457A
4-39. Equipment
Required.
A Frequency
Source capable of providing a 20 Hz (+ 016%)
and
a 1 MHz
(+ 003%) sine-wave signal is required for this procedure.
1. Set the HP 3457A to the Frequency Function (FREQ).
2. Set the Frequency Standard for a | volt, 20 Hz sine-wave output signal and connect it to the HI and
LO input terminals of the HP 3457A.
3. Use the Test Record to record the frequency readings at 20 Hz and | MHz.
4. If either of the readings are beyond
the limits specified, the instrument should
be calibrated
or
repaired, Refer to Section V of this manual for calibration procedures,
Rear Input Performance Tests (Standard Instrument)
OSENSE (4 WIRE)
HI
LO
HI
INPUT (2 WIRE)
0
450V pk
WAX
TO FRONT
TERMINALS
OR+
aii ty
©—a« CURRENT FUSE
REAR
TERMINALS
450V
pk MAX
3)
Figure 4-9. Standard Instrument Rear Input Performance Test Connections
4-40. Preliminary Steps
1, Short the Rear Input HI and LO terminals and the Q Sense HI and LO terminals as shown
in
Figure 4-9,
2. Select the Rear Terminal Input. (Press the TERM configuration key, enter the number 2 and press
the ENT key.
4-41. DC Voltage Function - Rear Terminal Offset Test
4-42. Equipment Required. A low thermal short (copper
re) is required for this procedure.
1. Set the HP 3457A to the DC Voltage function (DCV) and the number of digits displayed (DIGITS
DISP) to six.