HP 8110A Service Manual page 102

150 mhz pulse generator
Hide thumbs Also See for 8110A:
Table of Contents

Advertisement

Serial No .
Ambi ent temp erat ure
° C
Customer
R
elat ive humi dity
% ο
CSO#
Line
f
requency
Hz
T ested by
Date
Co mm ents :
Internal Oscillator Per i od
Scope Un certain ty factor
Test R esults fo r HP 8110 Α Mainframe
TR Entry
Test
L imi t
A
ctual
Limit
P
ass
Fail
Minimum
R
esult
Maximum
1 - 1
6 .65 ns 6 .2175 ns
7.0825 ns
1 - 2
9 .99 ns
9 .390 ns
10 .589 ns
1-3
10 .0 ns
9 .4 ns
10 .6 n s
1-4
50 .0 ns
47 .4 ns
52 .6
n
s
1 - 5
99 .9 ns 94 .805
n
s
104.995 ns
Co unter Un certainty factor
TR Entry
Test
Limit
Actua l
Limit
P
ass
Fail
Minimum
R
esult
Μαχίτηυm
1 - 6
100 ns
94 .9 ns
105 .1
n
s
1-7
500 ns
474.9 ns
525 .1
n
s
1-8
1 μs
949 .9 ns
1050 .1
n
s
Testing the HP 811 0Α 3-55

Advertisement

Table of Contents
loading

Table of Contents