Open the SBUS card retainers.
Holding the new D100 by the edges, unpack it. There are no jumpers to be set.
Holding the D100 at an angle, insert the card backplate behind the tabs of the SBUS
opening.
Lower the card into position. Carefully press the corners of the D100 to push it into the
connector. Do not force the card or you will damage the pins on the connector.
Push the card retainers forward over the edge of the SBUS card.
Fig. 12 Camera Interface D100
Switch on power to the system.
You may check the 5 V power on the test points (refer to Fig. 16).
Ensure that the system is operating properly.
MAMMOMAT 3000 Opdima