Upgrade of the MAMMOMAT software (generator and stand) to V2.2 (part No. 62 98 769, if the version to be upgraded is lower than V1.5, the AEC have to be upgraded, see Update Instruction, UI 001/96) Siemens-Elema AB Register 3 SPB7-230.033.10...
NOTE contains information provided with special emphasis to NOTE facilitate proper use of the equipment or proper execution of a procedure, i.e. hints, tips. Opdima? Register 3 SPB7-230.033.10 Page 2 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Automatic exposure control (AEC) The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdima? Measurement Certificates. Opdima? Register 3 SPB7-230.033.10 Page 4 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 9
Do not touch the pins in the contacts of the camera. The camera shall be used within 1030? C. The camera shall be transported or stored within 040? C. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 2...
Page 10
2 - 2 Protective measures for CCD camera This page intentionally left blank. Opdima? Register 3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 11
Risk of damaging components. Use ESD-equipment, ground prior to making contact and place the components on a conductive surface, see Technical Informa- tion TI 219. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 2 Opdima? Solna, Sweden Rev.
Page 12
3 - 2 Protective measures This page intentionally left blank. Opdima? Register 3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
*. Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions). Store the documentation and software CDs in a proper place for future service use. Opdima? Register 3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdima NOTE The workstation shall be placed at least 1.5 m from the object table of the stand. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 26 Opdima? Solna, Sweden Rev.
Remove the two screws (7/Fig. 2) at the top of the stand. Remove the cover (8/Fig. 2) by lifting it up. Fig. 2 Removal of covers Opdima? Register 3 SPB7-230.033.10 Page 2 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Pry loose the plastic strips (2/Fig. 3) on both sides using a screwdriver or similar tool. Remove the screws (3/Fig. 3) on both sides and remove the cover (4/Fig. 3). Fig. 3 Removal of covers Siemens-Elema AB Register 3 SPB7-230.033.10 Page 3 of 26 Opdima? Solna, Sweden Rev.
Install the two standoffs (4/Fig. 4) by using the screws (5/Fig. 4) and the washers (6/Fig. 4). 985 – 5 mm Fig. 4 Rear cover, from the back side Opdima? Register 3 SPB7-230.033.10 Page 4 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 19
Install the rear cover (2/Fig. 5) on the stand and attach the biopsy controller (1/Fig. 5) to the holders. Switch on the power-on switch on the biopsy controller. Fig. 5 Biopsy controller Siemens-Elema AB Register 3 SPB7-230.033.10 Page 5 of 26...
Install the new board D702 Fig. 7 in the generator. If upgrading from a version lower than V1.5, the AEC have to be NOTE upgraded, see Update Instruction, UI 001/96. Opdima? Register 3 SPB7-230.033.10 Page 6 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
If not already done in conjunction with installation of printer (option), an isolation board D707 (see Fig. 10) shall be installed. Install the board according to Installation Instructions for Isolation P.c. Board D707, RX B7-120.031.03..Siemens-Elema AB Register 3 SPB7-230.033.10...
Select Main menu => Configuration => Mischelaneous=> Cassette loaded switch and set the switch to ON. Switch off the MAMMOMAT with the main circuit breaker. Opdima? Register 3 SPB7-230.033.10 Page 8 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Z801 X880 Printer M3000 printer/ M3000 stand (option) Service PC X885 Biopsy unit M3000 generator (M3000 stand) Fig. 8 Connection of cables Siemens-Elema AB Register 3 SPB7-230.033.10 Page 9 of 26 Opdima? Solna, Sweden Rev. 03 08.02 SPS-UD Installation...
Use compartment according to the figure below. Serial number higher than 3000 Cable X206-X2 Serial number lower than 3000 Fig. 9 Cable-duct compartments Opdima? Register 3 SPB7-230.033.10 Page 10 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 25
This is to ensure electromagnetic compatibility of the equipment. Chassie ground clamp Fig. 10 Connection of cable to the generator Siemens-Elema AB Register 3 SPB7-230.033.10 Page 11 of 26 Opdima? Solna, Sweden Rev.
Page 26
The wall socket must be marked according to Fig. 24. Close the cable duct cover and install the front cover of the generator. Opdima? Register 3 SPB7-230.033.10 Page 12 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Remove the screws (1/Fig. 14) and remove the cover (2/Fig. 14) of the object table arm. Fig. 14 Object table arm before installation Opdima? Register 3 SPB7-230.033.10 Page 14 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 29
Install a new thread (1/Fig. 15). Fig. 15 Installation of thread Install the holder (1/Fig. 16) with the pull relief device, using included Allen screws (2/Fig. 16). Fig. 16 Object table arm after installation Siemens-Elema AB Register 3 SPB7-230.033.10 Page 15 of 26 Opdima? Solna, Sweden Rev.
Attach the supplied plastic strip around the opening of the cable outlet cover (4/Fig. 17). Install the cable outlet cover (4/Fig. 17). Mount the previously removed stand covers. Opdima? Register 3 SPB7-230.033.10 Page 16 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Connect the cable from the control unit (2/Fig. 18) for the vertically adjustment to the connector marked “Handschalter” (3/Fig. 18). Connect the cable from the motor (4/Fig. 18) for the vertically adjustment to the other connector (5/Fig. 18). Siemens-Elema AB Register 3 SPB7-230.033.10 Page 17 of 26...
Page 32
Fig. 19 System site overview NOTE There must be enough space around the workstation table so that it safely can be vertically adjusted . Opdima? Register 3 SPB7-230.033.10 Page 18 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 33
Place the main unit to the right on its shelf (1/Fig. 20). Mount the MO unit shelf close to the main unit (2/Fig. 20). Place the MO unit to the right on the shelf (2/Fig. 20). Siemens-Elema AB Register 3 SPB7-230.033.10...
-Place the plastic cover (2/Fig. 22) on the other side of the connectors. -Fasten the plastic screws (3/Fig. 22). Biopsy controller Fig. 22 Mounting the touch protection Opdima? Register 3 SPB7-230.033.10 Page 20 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 35
Work- station station Power On-Off breaker Power On-Off breaker Power connection Power connection in MAMMOMAT 3000-stand in MAMMOMAT 3000-stand Fig. 23 Power distribution Siemens-Elema AB Register 3 SPB7-230.033.10 Page 21 of 26 Opdima? Solna, Sweden Rev. 03 08.02 SPS-UD Installation...
Page 36
Connect the wall socket for the workstation and printer power supply to the same branch circuit as the MAMMOMAT power sup- ply. (The workstation table can be connected to any branch circuit.) Opdima? Register 3 SPB7-230.033.10 Page 22 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 37
10 Camera interface via biopsy controller (X203) Keyboard interface 11 Monitor interface 6A Mammomat interface via biopsy controller (X204) 6B Biopsy controller interface (X204) Fig. 25 Workstation main unit connectors Siemens-Elema AB Register 3 SPB7-230.033.10 Page 23 of 26 Opdima? Solna, Sweden Rev.
Connect one end of the network cable (included in the delivery) to the workstation. Connect the other end of the network cable to a network socket. Log in with the user name “service”. The password is obtained from Siemens Service Centre.
Page 39
Add access to the network printer in the advanced service dialog, see Networked Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions. Select printer in the advanced service dialog, see Networked Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 25 of 26 Opdima? Solna, Sweden Rev.
Before the installed workstation table (option) is handed over to the customer, check that the workstation table can be vertically adjusted between the highest and lowest possible positions properly and safety. Opdima? Register 3 SPB7-230.033.10 Page 26 of 26 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Filling in of LINA CARD, see Page 6 - 2 • Performing of DHHS maintenance, see Page 6 - 2 The Opdima? external diaphragm must be used for tests and cali- NOTICE brations involving radiation. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 1 of 4 Opdima? Solna, Sweden Rev.
The workstation is started according to instructions in Starting up the workstation and login in Supplement to the Instructions for Use MAMMOMAT 3000 - Opdima?. Use the user name “service” and the password obtained from Siemens service center when entering the system.
Page 43
MAMMOMAT. Therefore the DHHS Maintenance Instructions supplied with the MAMMOMAT can be removed. However, the Measurement Certificates supplied with the MAMMOMAT should be kept for future reference, if it contains previous measurement results. Siemens-Elema AB Register 3 SPB7-230.033.10 Page 3 of 4 Opdima? Solna, Sweden Rev.
Page 44
6 - 4 Final procedures This page intentionally left blank. Opdima? Register 3 SPB7-230.033.10 Page 4 of 4 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 45
Text change in Caution. Notice G Note, Notice G Caution (text change). Page 5 - 24 Notice G Note. Page 5 - 25 Notice G Note. Page 6 - 3 Siemens-Elema AB Register_3 SPB7-230.033.10 Page 1 of 2 Opdima? Solna, Sweden Rev. 03 08.02...
Page 46
7 - 2 Changes to previous version This page intentionally left blank. Opdima? Register_3 SPB7-230.033.10 Page 2 of 2 Siemens-Elema AB Installation Rev. 03 08.02 SPS-UD Solna, Sweden...
Page 47
150 mm Drilling template A Drill five holes with a 5 mm drill according to template A and B. Place template B underneath template A with no space between them. Check distances before drilling!